beautypg.com

4 error messages, 4 messages d’erreur, 17 english français – Xylem TTX 100/110/120 User Manual

Page 17

background image

17

English

Français

4 Error messages

The following error messages are displayed
while using the TTX thermometer.

“Hi“ --> for temperature to be measured above
+350°C/+1200°C.

“Lo“ --> for temperature to be measured under
-50°C/-60°C.

“Er 2“ --> too quick a change in the ambient
temperature.

“Er 3“ --> Ambient temperature either
-20°C or above +50°C.

--> Bring the device within the prescribed
operating temperature range and wait for about
30 min.

“Er “ --> for all other probable errors.

--> “Reset“ to be executed after removing the
battery from the battery compartment, (see
point 6.3)
wait for about 1 min, insert the battery
and restart the device.

4 Messages d’erreur

Pendant l´utilisation du thermomètre TTX, les
messages d´erreur suivants peuvent s´affi cher
sur le display.

« Hi » --> la température à mesurer se situe
au-delà de +350°C/+1200°C.

« Lo » --> la temperature à mesurer se situe
au-deçà de -50°C/-60°C.

« Er 2 » --> changement trop rapide de la tem-
pérature ambiante.

« Er 3 » --> la température ambiante est sous
-20°C ou au delà de +55°C.

-->Recréez les conditions atmosphériques
prescrites pour l´usage de l´appareil et attendez
30 min.

« Er » --> pour toutes les autres erreurs possib-
les.

--> exécuter la réinitialisation ou « Reset » pen-
dant que vous ôtez la batterie du boîtier, ( voir
le point 6.3)
attendez environ 1 min. insérez de
nouveau la batterie et enclenchez l´appareil.