Info messages
EI1331
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components
EI2331
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components,
replace components
ES1331
Component hardware defective
* Contact service
ESB331
SensReg: Electrolyte supply is depleted
* Change electrolyte solution and membrane head immediately (see
operating manual)
ESC331
SensLeck: Membrane head damaged
* Change electrolyte solution and membrane head immediately (see
operating manual)
Note: This piece of information is transmitted with the
TriOxmatic
®
700 IQ only.
Message code
Message text
Message code
Message text
IC1331
Sensor has been successfully calibrated
* For calibration data, see calibration history
IC4331
Last valid calibration has been activated. Make sure the sensor
operates correctly.
II1331
Language not available,
Default language German
* Contact service