3 calibration history, 4 reactivation of the last valid calibration, Calibration history -4 – YSI IQ S ENSOR N ET TriOxmatic 700 IQ (SW) User Manual
Page 20: Reactivation of the last valid calibration -4
Measuring / Operation
TriOxmatic
®
700 IQ (SW)
4 - 4
ba76013e01
01/2012
4.2.3
Calibration history
Calibration history
(available in the
IQ S
ENSOR
N
ET
system 2020 XT only)
Fig. 4-1
Calibration history TriOxmatic
®
700 IQ (SW)
The calibration history contains the following information:
Date of the calibration
Relative Slope (dimensionless)
Rating of the calibration:
– o.k.: Calibration
was
successful.
The new calibration values are used for the measurements.
– Error: Calibration was not successful.
The Sensor is blocked for measurement.
4.2.4
Reactivation of the last valid calibration
The TriOxmatic
®
700 IQ (SW) enables you to reactivate the last valid
calibration when needed. Thereby, it is possible to continue with
measurements, when a calibration failed or it is assumed that optimal
calibration conditions were not met.
Note
The reactivation of former calibration data is only a temporary solution.
Please consider, that thereby possibly wrong measurement values are
produced. Please check correct functioning of the sensor by performing
a function check and/ or a user calibration.
Reactivating the last
valid calibration data
actual active
calibration
chronological list
with the last
calibrations
1
Open the the setting table (see section 3.4).
2
Choose in the menue item Calibration the setting last valid and
quit the setting table afterwards with Save and quit.