3
Commissioning
3.1
Scope of delivery
TriOxmatic
®
700 IQ (SW)
ZBK 600 accessory kit
The sensor is filled with electrolyte and provided with a protective
hood
and protective caps.
Operating manual.
3.2
Installation
Connection cable
A sensor connection cable of the SACIQ or SACIQ SW type is required
to connect the sensor. The cable is available in different lengths.
Compared to the standard model SACIQ, the SACIQ SW sensor
connection cable is optimized regarding its corrosion resistance in
seawater and brackish water and adapted for use in conjunction with
the TriOxmatic
®
700 IQ SW. Information on this and other
IQ S
ENSOR
N
ET
accessories is given in the YSI catalog and on the
Internet.
Note
How to connect the SACIQ (SW) sensor connection cable to the
terminal strip of an MIQ module is described in chapter 3 I
NSTALLATION
of the IQ S
ENSOR
N
ET
system operating manual.
Are the plug
connections dry?
Before connecting the sensor and sensor connection cable, please
make sure that the plug connections are dry. If moisture gets into the
plug connections, first dry the plug connections (dab them dry or blow
them dry using compressed air).
Note
Do not suspend the sensor on the sensor connection cable. Use a
sensor holder or an armature. Information on this and other
IQ S
ENSOR
N
ET
accessories is given in the YSI catalog and on the
Internet.
Minimum approach flow
The minimum required approach flow at the sensor must be present
during measurement (see chapter 7 T
). The minimum
approach flow can, e.g. be present due to:
natural movement of the water (flow rate, sea disturbance)
turbulence in the activated sludge basin.