Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
700 IQ (SW)
5 - 2
ba76013e01
01/2012
5.2
Cleaning the sensor shaft and membrane
For normal operation (e.g. municipal wastewater), cleaning and
calibration are strongly recommended:
in the case of pollution (according to visual check)
in the case of suspected incorrect measured values (usually too low)
if the measured value lies outside the range of precision required by
the user during the function check.
Cleaning agents
Cleaning
Caution
Acetic acid irritates the eyes and the skin. When handling acetic
acid, always wear protective gloves and protective goggles.
Contamination
Cleaning agents
For sludge and loosely
adhering dirt, or biological
deposits
Soft cloth or soft sponge, warm
tapwater with detergent
Salt and / or lime deposits
Acetic acid (volume percentage =
20 %),
soft cloth or soft sponge
1
Pull the sensor out of the sample.
2
Get rid of any coarse contamination on the sensor (e.g. brush
it off in a bucket of tapwater, wash it down with a hose or wash
it off with a cloth).
3
Unscrew the protective hood from the sensor.