9.1
Explanation of the messages
This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S
ENSOR
N
ET
system for the TriOxmatic
®
700 IQ (SW) sensor.
Note
Information on the contents and structure of the log book, and how to
call it up, is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system operating manual.
Note
All message codes of the TriOxmatic
®
700 IQ (SW) end up with "331".
9.1.1
Error messages
Message code
Message text
EA1331
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA2331
Sensor temperature too high!
* Check process and application
EA3331
Sensor temperature too low!
* Check process and application
EC4331
Sensor could not be calibrated,
sensor blocked for measurement
Cause: instable signal
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)
EC5331
Sensor could not be calibrated,
sensor blocked for measurement
Cause: slope error
* Check polarization time and temperature adjustment
* Check calibration conditions
* Read calibration history
* Repeat calibration, if unsuccessful: service the sensor (see operating
manual)