YSI IQ S ENSOR N ET TriOxmatic 700 IQ (SW) User Manual
Page 42
Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
700 IQ (SW)
5 - 20
ba76013e01
01/2012
5.5
Checking the sensor for freedom from zero-current
The sensor is zero-current free. Therefore, calibration of the sensor in
oxygen-free solution is not provided.
However, it is possible to check the freedom from zero-current
according to DIN EN 25814/ISO 5814. This is advisable if there is any
suspicion of a malfunction.
Test solution
1 g/l aqueous sodium sulfite solution, Na
2
SO
3
(addition of 1 mg/l of a cobalt(II) salt accelerates the removal of oxygen
from the solution.)
The sensor should be in operation for at least 1 hour before the
inspection.
Test procedure
Test criterion
The sensor is OK if < 2 % oxygen saturation is displayed as the
measured value.
If the measured value is > 2 %:
Clean the gold working electrode (see section 5.4.1 C
) and change the electrolyte and
membrane cap (see section 5.3 C
).
If necessary, also clean the silver counter electrode (see section
).
1
Clean the sensor (see section 5.2 C
2
In the sensor setting table, select the measuring mode
Saturation (unit %) and switch to the measured value display
with
m
.
3
Submerse the sensor in the test solution.
4
Leave the sensor in the test solution for an hour.
5
Read the measured value.