4 electrical data, 5 characteristic data on delivery, Electrical data -4 – YSI IQ S ENSOR N ET TriOxmatic 700 IQ (SW) User Manual
Page 50: Characteristic data on delivery -4
Technical data
TriOxmatic
®
700 IQ (SW)
7 - 4
ba76013e01
01/2012
Automatic sensor
monitoring
(SensCheck function)
* Note: The leak recognition can be selected in the setting table of the
Sensor monitoring relay function but the function is not active.
Instrument safety
7.4
Electrical data
7.5
Characteristic data on delivery
Zero signal
< 0.2 % of the saturation value
Response time at 25 °C
t
90
(90 % of the final value display after) < 180 s
Own consumption
0.0059 µg h
-1
(mg/l)
-1
at 20 °C
Drift
< 1 % per month during permanent polarization
Working life per
electrolyte filling
TriOxmatic
®
700 IQ
TriOxmatic
®
700 IQ SW
SensReg
(electrolyte solution depleted)
Yes
Yes
SensLeck
(membrane cap leaking)
Yes
No *
Applicable norms
– EN 61010-1
– UL 3111-1
– CAN/CSA C22.2 No. 1010.1
Nominal voltage
Max. 24VDC
via the IQ S
ENSOR
N
ET
(for more
details, see chapter T
ECHNICAL
DATA
of the IQ S
ENSOR
N
ET
system operating manual)
Power consumption
0.2 W
Protective class
III
D. O. concentration
Theoretical electrolyte reserve
Air saturation
approx. 5 years
< 2 mg/l
approx. 25 years