Installing the interface in the
IQ S
ENSOR
N
ET
2.1
Installing the RS485 interface
Profibus and Modbus use the same interface (RS485) for data
exchange.
We especially recommend the "Aufbaurichtlinie PROFIBUS/FMS",
published by the PROFIBUS user organization PNO, (PNO order num-
ber 2.111) with many practical tips for installation that also apply to
Modbus systems (
http://www.profibus.com/download/installa-
tion-guide/
).
Further instructions for the installation and operation of a Modbus net-
work are availabe on the Internet under www.modbus.org.
Detailed information on the subject of PROFIBUS is given on the Inter-
net by the PROFIBUS user organization (PNO) under www.profi-
bus.com.
Connecting the Profibus
/ Modbus cable
The Profibus / Modbus cable is connected to the IQ S
ENSOR
N
ET
using
a 9-pin D-SUB connector on the top of the housing.
Suitable cables
Connectors to be used
To remove the blind plug, lift the safety bracket with a suitable screw
driver at both clamps on the left and right side (see Fig. 2-1).
Field bus type
Cable
Profibus
PROFIBUS cable
Modbus
Cable for RS485 (e.g. PROFIBUS cable)
Manufacturer
Type
Phoenix Contact GmbH &
Co. KG
Flachsmarkt 8
32825 Blomberg
Germany
http://www.phoenixcon-
tact.com
VS-09-PROFB-SC
(Phoenix article no. 1654549, with
screwed contacts,
available from YSI under the order no.
902 888Y)
VS-09-PROFB-SP
(Phoenix article no. 1654345, with spring
contacts)