Specifi cations – Xylem HandyLab 780 User Manual
Page 80
![background image](/manuals/310616/80/background.png)
80
Specifi cations
Display
User interface
Straightforward menu navigation with graphic icons and detailed operating
instructions in plain text
Sensoface
Gives you indications about the status of the sensor, analysis of zero-point/
slope, setting time, frequency of calibration. Status indication (friendly, neutral,
sad)
Grafi k Display
QVGA TFT display with white backlighting
Status indicators
For battery power level, logger
Hinweise
Hourglass
Tastatur
[on/off], [meas], [enter], [], [], [], [], 2 context-sensitive softkeys
Diagnostics functions
Sensor data
(Memosens only)
Manufacturer, sensor type, serial number,
operating time
Calibration data
Calibration date, zero, slope
Device self-test
Automatic memory test (FLASH, EEPROM, RAM)
Device data
Device type, software version, hardware version
Data retention
Parameters, calibration data > 10 years
Data logger
1x Micro USB-B for data transfer to PC
Data logger
10.000 Speicherplätze
Aufzeichnung:
manuell, intervall- oder ereignisgesteuert
Cal data logger
MemoLog
(Memosens only)
Up to 100 Memosens calibration records can be saved
– Recording viewable on the display
– Directly retrievable via MemoSuite (USB)
Manufacturer, sensor type, serial no., zero, slope, calibration date
Communication
USB 2.0
Profi le
HID, installation without drive
Use
Data exchange and confi guration via HandLab
Pilot software
RoHS conformity
According to directive 2011/65/EC
Nominal operating
conditions
Ambient temperature
–10 … +55 °C
Transport/ Storage
temperature
–25 … +70 °C
Relative humidity
0 … 95 %, short-term condensing allowed
Power
Power supply
Batterys 4x AA (Mignon), 4x Akku NiMH or
1x Li-Ionen-Akku,
rechargeable via USB
Operating time
approx 500 h (Alkaline)
Housing
Material
PA12 GF30 + TPE
Protection
IP66/67 with pressure compensation
Dimensions
approx. (132 x 156 x 30) mm
Weight
approx. 500 g
Licensing - Compliance - Equipment safety
EMC
EN 61326-1 (General Requirements)
Emitted interference
Class B (residential area)
Immunity to interference
Industry
EN 61326-2-3 (Particular Requirements for Transmitters)