Specifi cations – Xylem HandyLab 780 User Manual
Page 79
79
Specifi cations
pH-calibration
Operating modes
*)
AutoCal
Calibration with automatic buffer recognition
MANUAL
Manual calibration with entry of individual buffer values
DATA INPUT
Data entry of zero and slope
-01-
(Default)
NIST standard
1.679/4.006/6.865/9.180
-02-
NIST technical
1.68/4.00/7.00/10.01/12.46
-03-
SI Analytics technical
2.00/4.00/7.00/10.00
-04-
Knick Calimat
2.00/4.00/7.00/9.00/12.00
-05-
Mettler Toledo
2.00/4.01/7.00/9.21
-06-
HACH
4.01/7.00/10.01/12.00
-07-
Ciba (94)
2.06/4.00/7.00/10.00
-08-
Hamilton
2.00/4.01/7.00/10.01/12.00
-09-
Reagecon
2.00/4.00/7.00/9.00/12.00
-10-
DIN 19267
1.09/4.65/6.79/9.23/12.75
-U1-
via HandyLab Pilot (User)
Permissible calibration
range
Zero point
pH 6 … 8
With ISFET:
–750 … +750 mV Operating point (asymmetry)
Slope
approx. 74 … 104 %
Calibration timer
*)
Interval 1 … 99 days, can be switched off
Memosens input, oxygen
M8 socket, 4 pins, for Memosens lab cable
Display ranges
1)
Saturation
0.000 ... 200.0 %
Concentration
000 μg/l … 20.00 mg/l
Partial pressure
0.0 … 1000 mbars
Temperature meas. range
1)
-20 … 150 °C
Sensor standardization:
Automatic calibration in air, humidity adjustable,
Zero calibration
AutoCal buffer sets
*)
*) User-defi ned
1) Ranges depending on Memosens sensor
2) According to EN 60746-1, at nominal operating conditions
3) ± 1 count
4) Plus sensor error