Specifi cations – Xylem HandyLab 780 User Manual

Page 79

background image

79

Specifi cations

pH-calibration
Operating modes

*)

AutoCal

Calibration with automatic buffer recognition

MANUAL

Manual calibration with entry of individual buffer values

DATA INPUT

Data entry of zero and slope

-01-
(Default)

NIST standard

1.679/4.006/6.865/9.180

-02-

NIST technical

1.68/4.00/7.00/10.01/12.46

-03-

SI Analytics technical

2.00/4.00/7.00/10.00

-04-

Knick Calimat

2.00/4.00/7.00/9.00/12.00

-05-

Mettler Toledo

2.00/4.01/7.00/9.21

-06-

HACH

4.01/7.00/10.01/12.00

-07-

Ciba (94)

2.06/4.00/7.00/10.00

-08-

Hamilton

2.00/4.01/7.00/10.01/12.00

-09-

Reagecon

2.00/4.00/7.00/9.00/12.00

-10-

DIN 19267

1.09/4.65/6.79/9.23/12.75

-U1-

via HandyLab Pilot (User)

Permissible calibration
range

Zero point

pH 6 … 8

With ISFET:

–750 … +750 mV Operating point (asymmetry)

Slope

approx. 74 … 104 %

Calibration timer

*)

Interval 1 … 99 days, can be switched off

Memosens input, oxygen

M8 socket, 4 pins, for Memosens lab cable

Display ranges

1)

Saturation

0.000 ... 200.0 %

Concentration

000 μg/l … 20.00 mg/l

Partial pressure

0.0 … 1000 mbars

Temperature meas. range

1)

-20 … 150 °C

Sensor standardization:

Automatic calibration in air, humidity adjustable,
Zero calibration

AutoCal buffer sets

*)

*) User-defi ned
1) Ranges depending on Memosens sensor
2) According to EN 60746-1, at nominal operating conditions
3) ± 1 count
4) Plus sensor error