beautypg.com

Confi guration, Cond, Conductivity confi guration – Xylem HandyLab 780 User Manual

Page 56

background image

56

Cond

Confi guration

Conductivity Confi guration

1) In measuring mode, press the Menu softkey.
2) Select “Confi guration” and confi rm by pressing enter.
3) Make the desired adjustments.
The following table gives you an overview. Factory settings are shown in bold
print

.

Language

Deutsch

| English | Español | Italiano | Fran-

çais |

Português

Auto-off

Off

| 5 min | 10 min | 30 min | 60 min

Temperature

°C

| °F

+ Cond sensor

*)

Conductivity

S/cm

| S/m

Range selec-
tion

Auto

| o.ooo μS/cm | oo.oo μS/cm |

ooo.o μS/cm | oooo μS/cm |
oo.oo mS/cm | ooo.o mS/cm | oooo mS/cm

Calculation

Off

| MΩ cm | TC | SAL | TDS | % by wt

TC comp.

TC: Linear | NLF | NaCl | HCl | NH3 | NaOH

TC of solution

TC: 0 … 20.0 %/K | 1.00 %/K

Ref. temp

TC: 0 … 100.0 °C | 25 °C 32 … 212 °F |
77 °F

TDS factor

TDS: 0 …9.99 | 1.00

Solution

% by wt: NaCl | HCl | NaOH | H2SO4 | HNO3

“Confi guration” menu selection – part 1

*) “+” indicates that submenus can be opened by pressing enter.