beautypg.com

Calibration, Cond, Entry of solution” calibration – Xylem HandyLab 780 User Manual

Page 64: Cell constant calibration

background image

64

Cond

Calibration

“Entry of Solution” Calibration

(Calibration by entering the conductivity with display of the cell con-
stant)

1) Press the Start softkey.
2) Immerse the sensor in the solution.
3) Enter the temperature-corrected conductivity value and press enter.
4) Finally, you can Apply these values or Cancel the calibration.

Cell Constant Calibration

(Calibration by entering the cell constant with display of the conductivi-
ty)

1) Press the Start softkey.
2) Immerse the sensor in the solution.
3) Modify the value of the cell constant until the temperature-corrected

conductivity value is reached. Then press enter.

4) Finally, you can Apply these values or Cancel the calibration.

Note:

To abort calibration, you can press meas at any time.