Information, Ph oxy cond, Sensor monitor – Xylem HandyLab 780 User Manual
Page 52
![background image](/manuals/310616/52/background.png)
52
pH
Oxy Cond
Information
MemoLog (Memosens only)
Displays the individual calibration records. You have the possibility to delete
individual entries or all entries. The following parameters are displayed:
• Sensor type
• Manufacturer
• Calibration date
• Serial no.
• Zero point
• Slope
• Load conditions
• Measuring point (TAG)
Background
: The device provides a calibration data logger, which must be
activated in the confi guration menu. With “MemoLog” activated, up to 100
calibration records can be directly saved to the device. Then, the complete
Memosens index structure will be recorded after every calibration. The Me-
moSuite software allows convenient management of the calibration data.
Sensor Monitor
Shows the raw values available fom the connected sensor:
pH, analog
mV, temperature, temperature detector, temp. resis-
tance
pH, digital, glass
mV, temperature, glass impedance
pH, digital, ISFET
mV, leakage current, temperature
pH, ORP
mV, temperature
Cond, analog
Resistance, conductance, temperature, temperature
detector, temperature resistance
Cond, digital
Resistance, conductance, temperature
Oxy, digital
Sensor current, leakage current, polarization voltage,
partial pressure, air pressure, temperature