Measuring, Ph oxy cond, Switching the measured value display – Xylem HandyLab 780 User Manual
Page 67: Adjusting the temperature

67
pH
Oxy Cond
Measuring
Measuring
Once you have completed all preparations, you can start with the actual
measurement.
1) Connect the desired sensor to the meter. Some sensors require a special
preparation. Please proceed according to the operating instructions for
the sensor.
2) Switch the meter on using the on/off or meas key.
3) Depending on the measurement method and the sensor used, immerse
the sensing part of the sensor in the medium to be measured.
4) Watch the display and wait for the reading to stabilize.
Note
: Measurement can also be controlled via the HandyLab Pilot software.
Switching the Measured Value Display
During measurement, you can toggle between display of
primary / secondary measured value and clock by pressing meas.
Adjusting the Temperature
When you connect a sensor without temperature detector, you can manually
adjust the temperature for measurement or calibration:
1) Press meas to access measuring mode.
The adjusted temperature will be displayed.
2) Set the desired temperature value using the or arrow.
Holding the key depressed changes the temperature value at high speed.