Appendix C

NIR256-2.5µm Sensitivity

When configuring a system for operation out to 2.5µ

m it is important that you consider the following

details:

• Short Integration Times: For the detector to be sensitive out to 2.5µm, the detector’s band gap

energy must be small. Unfortunately, this raises the absolute level of the detectors dark signal. A
typical dark signal at 34ms is shown below. While the detector functions at integration times
greater than 50ms, this seems to be the realistic limit.

• Fiber Selection: For maximum signal intensity, alternative fiber materials should be used for

wavelengths greater than 2.2µ

m.

The figure below shows a dark signal at 24ms integration time. Full Scale is 65,535 counts. The spectra
shows the presence of a defective pixel.