3: Troubleshooting
3. Delete
the
OOI_USB.INF in the INF directory. If your computer is running either the Windows
2000 or XP operating system, you must also delete the OOI_USB.PNF file in the INF directory.
4. Navigate to the Windows | System32 | Drivers directory.
5. Delete
the
EZUSB.SYS file.
6. Reinstall your Ocean Optics application and reboot the system when prompted.
7. Plug in the USB device.
The system is now able to locate and install the correct drivers for the USB device.
Mac Operating Systems
Since there are no device files for the NIR Spectrometer in a Mac operating system, you should not
encounter any problems if you installed the spectrometer before the SpectraSuite software.
Linux Operating Systems
For Linux operating systems, all you need to do is install the SpectraSuite software, then unplug and
replug in the spectrometer. Technically, the driver files for Linux simply give nonprivileged users
permission to use newly connected hardware. There isn’t any long-term harm to plugging in the device
before installing the software.
Troubleshooting the Serial Port Configuration
Occasionally, you may encounter problems with the serial port connection and/or software. Perform the
following procedure to troubleshoot the serial port connection.
►
Procedure
1. Cycle the power on the NIR Spectrometer and restart the OOIBase32 software. This ensures that
the software and the hardware synchronize properly.
2. Determine the serial port (COM port) number:
Operating System Instructions
Windows
95/98/ME/XP
1. Right-click on My Computer
2. Select Properties
3. Click the Device Manager tab