A: Calibrating the Wavelength of the Spectrometer
7. Record the Intercept as well as the First, Second, and Third Coefficients. Also, look at the value
for R squared. It should be very close to one. If it is not, you have probably assigned one of your
wavelengths incorrectly. Keep these values at hand.
Saving the New Calibration Coefficients: USB
Mode
Wavelength calibration coefficients unique to each NIR Spectrometer are programmed into an EEPROM
memory chip on the NIR Spectrometer.
You can save over old calibration coefficients with new ones and the spectrometer operating software can
read these coefficients, but only if you are using the NIR Spectrometer via the USB port. If you are using
the NIR Spectrometer via the serial port, see the
Saving the New Calibration Coefficients: Serial Mode
section later in this Appendix.
►
Procedure
To Save Wavelength Calibration Coefficients Using the USB Mode
1. Ensure that the NIR Spectrometer is connected to the PC and that no other applications are
running.
2. Point your browser to
down to Microcode. Double-click on USB EEPROM Programmer.
3. Save the setup file to your computer.
4. Run
the
Setup.exe file to install the software
5. At
the
Welcome screen, click the Next button.
6. At
the
Destination Location screen, accept the default or click the Browse button to pick a
directory. Then, click the Next button.
7. Select a Program Manager Group and click the Next button. The Start Installation screen appears.
Click the Next button.
8. Click
the
Finish button when the Installation Complete screen appears.
9. Reboot the computer when prompted.
10. Navigate to USB EEPROM Programmer and open the software.
11. Click on the NIR Spectrometer device, located in the left pane of the USB Programmer screen.
12. Double-click on each of the calibration coefficients displayed in the right pane of the USB
Programmer screen and enter the new values acquired in Steps 5 and 6 of the
Wavelength of the Spectrometer
section in this Appendix.