External Triggering Options Instructions

6

200-00000-000-01-201401

Product

Supported

Not Supported

OOISensors

External Software

External Synchronization and External
Hardware

OOIIrrad

External Software and External Hardware

External Synchronization

Palm Spec

External Software

External Sync and External Hardware

Configuring External Triggering

External Software Trigger Mode

In this level-triggered mode, the spectrometer is “free running,” just as it is in the Normal mode. The
spectrometer is continually scanning and collecting data. With each trigger, the data collected up to the
trigger event is transferred to the software. If you continuously apply triggers (for example, by holding
down the button on via an external switch), this mode is equivalent to operating in the Normal mode.

In the External Software Trigger mode, you set the integration time (as well as all other acquisition
parameters) in the software. The source for the integration clock comes from the A/D converter.

If the software trigger is asserted during integration cycle n, the photons from this integration period will be read out and digitized at the start of

integration cycle n+1

External Software Triggering – Trigger Timing

When to Use the External Software Trigger Mode

Use the External Software Trigger mode if you:

• Are using a continuous illumination source, and the light intensity is constant before, during, and

after the trigger.

• Need to set the integration time in the software.