External Triggering Instructions
200-00000-000-01-201401
5
Integration Times Available with External
Triggering
The following table illustrates the integration times available with External Triggering modes:
Spectrometer
External
Software
External
Hardware
External
Synchronization
USB2000/HR2000
Set in Software
50 msec
Cannot Manually
Set
USB4000/USB2000+/Jaz/HR2000+/HR4000 Set in Software
Set in Software
Input by software
NIR Series
Set in Software
Not Available
Not Available
NIRQuest Series
Set in Software
Not Available
Not Available
S2000/S1024DW
Set in Software
Set on Jumpers
Cannot Manually
Set
S2000 D-Series
Set in Software
55 msec
Cannot Manually
Set
QE65000
Set in Software
Not Available
Not Available
Software Support
The following table details External Triggering support capacity in all Ocean Optics software products:
Product
Supported
Not Supported
SpectraSuite
All Modes
--
OOIBase
All Modes
--
OOIBase32
All Modes
--
OOIBase32 Platinum
All Modes
--
OOIColor
External Software and External Hardware
External Synchronization
OOI Driver Products
All Modes
--