Ocean Optics MultiFrequency Phase Fluorometer User Manual
Page 27
![background image](/manuals/292635/27/background.png)
3: Configuration
6. Select the Color assigned to each channel.
7. Select one of the following for Calibration Order. This selection defines the required
calibration.
• Linear (Stern Volmer) – Requires a minimum of two oxygen standards for calibration.
• Second Order Polynomial – Requires a minimum of three oxygen standards for calibration.
8. Select the Oxygen Units.
9. If desired, check the Temperature checkbox. If you select this item, you need either a factory
multi-temperature calibration or you must perform one yourself. Once you have stored a multi-
temperature calibration in OOISensors, a single-temperature calibration will be required.
10. If desired, check the Pressure Compensation checkbox. If you select this item, you need to have
a pressure transducer that is supported by the hardware and software.
11. For Henry’s coefficients, select either User (user-specified), or the given Water values.
12. If there is more than one MFPF channel, fill in the same information for Channel 2.
13. Click OK to save your MFPF configuration.
MFPF-00000-000-02-1207