3: Configuration

2. Set the MFPF parameters (located in the Instrument Controls box on the lower left of the main

screen) according to the following recommendations:

Parameter Setting

Dual Channel

Unchecked for set up, checked if 2 sensors are used

Sampling Period

2 seconds for set up

Frequency kHz

5 kHz – FOSPOR
40 kHz – FOXY and HIOXY

APD voltage

110 to 155 V photon-to-electron ratio

LED Channel

Depends on your experiment

Detector Channel

Depends on your experiment

3. Select Advanced | Show More Control Panel from the menu bar. The following dialog box

appears:

4. Set the parameters according to the following recommendations:

Parameter Setting

Analog AGC

Checked

Digital AGC

Checked

LED Mode

Intermittent

LED Percentage

25 -- 100%

16

MFPF-00000-000-02-1207