Troubleshooting, Deuterium lamp, Chapter 4 – Ocean Optics DH-2000 User Manual

Page 21

background image

000-10000-025-02-1209

13

Chapter 4

Troubleshooting

Deuterium Lamp

If the power supply or Deuterium lamp does not seem to functioning properly, check the following:

Issue

Probable Cause

Resolution

Line power not present

Check line voltage

Power switches on, but no
LEDs light.

Fuse defective

Check fuse

Deuterium lamp too hot

Allow the Deuterium lamp to cool
down (20 minutes).

Press On/Off switch again to
reset the Deuterium Tungsten-
Halogen lamp, then press again
to restart.

Deuterium lamp life exhausted

Replace Deuterium lamp

Deuterium lamp does not light.

The two-color LED under the
Deuterium On/Off switch lights
up red, indicating an error.

Deuterium lamp’s internal
connection plug is not closed
correctly.

Open unit (see the

Maintenance

)

and close connector plug.

Deuterium lamp turns off
during operation.

Deuterium lamp too hot

Turn off the unit.

Allow the unit to cool down for at
least 20 minutes.

Once the unit has cooled down,
turn the Deuterium lamp back on.