beautypg.com

Rainbow Electronics MAX188 User Manual

Page 15

background image

Power-Down

Choosing Power-Down Mode

You can save power by placing the converter in a
low-current shutdown state between conversions.
Select full power-down or fast power-down mode via
bits 7 and 8 of the DIN control byte with

SHDN high or

floating (see Tables 2 and 6). Pull

SHDN low at any time

to shut down the converter completely.

SHDN overrides

bits 7 and 8 of DIN word (see Table 7).

Full power-down mode turns off all chip functions that draw
quiescent current, reducing I

DD

and I

SS

typically to 2µA.

Fast power-down mode turns off all circuitry except the
bandgap reference. With the fast power-down mode, the
supply current is 30µA. Power-up time can be shortened
to 5µs in internal compensation mode.

In both software shutdown modes, the serial interface
remains operational, however, the ADC will not convert.
Table 5 illustrates how the choice of reference-buffer
compensation and power-down mode affects both
power-up delay and maximum sample rate.

In external compensation mode, the power-up time is
20ms with a 4.7µF compensation capacitor (200ms with
a 33µF capacitor) when the capacitor is fully discharged.
In fast power-down, you can eliminate start-up time by

using low-leakage capacitors that will not discharge
more than 1/2LSB while shut down. In shutdown, the
capacitor has to supply the current into the reference
(1.5µA typ) and the transient currents at power-up.

Figures 12a and 12b illustrate the various power-down
sequences in both external and internal clock modes.

Software Power-Down

Software power-down is activated using bits PD1 and
PD0 of the control byte. As shown in Table 6, PD1 and
PD0 also specify the clock mode. When software shut-
down is asserted, the ADC will continue to operate in
the last specified clock mode until the conversion is
complete. Then the ADC powers down into a low quies-
cent-current state. In internal clock mode, the interface
remains active and conversion results may be clocked
out while the MAX186/MAX188 have already entered a
software power-down.

The first logical 1 on DIN will be interpreted as a start
bit, and powers up the MAX186/MAX188. Following the
start bit, the data input word or control byte also deter-
mines clock and power-down modes. For example, if
the DIN word contains PD1 = 1, then the chip will
remain powered up. If PD1 = 0, a power-down will
resume after one conversion.

MAX186/MAX188

Low-Power, 8-Channel,

Serial 12-Bit ADCs

______________________________________________________________________________________

15

SCLK

DIN

DOUT

CS

S

CONTROL BYTE 0

CONTROL BYTE 1

S

CONVERSION RESULT 0

B11 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1 B0

CONVERSION RESULT 1

SSTRB

B11 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1 B0

CONTROL BYTE 2

S

1

8

1

8

1

CS

SCLK

DIN

DOUT

S

CONTROL BYTE 0

CONTROL BYTE 1

S

CONVERSION RESULT 0

B11 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1 B0

B11 B10 B9 B8

CONVERSION RESULT 1

• • •

• • •

• • •

• • •

Figure 11a. External Clock Mode, 15 Clocks/Conversion Timing

Figure 11b. External Clock Mode, 16 Clocks/Conversion Timing