Measurement and power solutions, Model 2220 specifications – Trek 2220 Piezo Driver-Power Amplifier User Manual
Page 2

Features
Digital Enable
A BNC connection for a TTL compatible signal
to turn ON/OFF the high voltage output is
provided. TTL high (or open) turns off the HV
output; TTL low turns on the HV output.
Response
A graduated 1-turn panel potentiometer is used
to optimize the AC response for various load
parameters.
High Voltage LED
Front panel red LED illuminates when the high
voltage is on.
Mechanical
Dimensions
85 mm H x 205 mm W 325 mm D
(3.3” H x 8.1” W x 12.8” D)
Weight
2 kg (4.4 lb)
HV Connector
SHV Connector
BNC Connectors
Amplifier Input, Voltage Monitor, Current Monitor,
Digital Enable
Operating Conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative Humidity
To 85%, noncondensing
Altitude
To 2000 meters (6561.68 ft.)
Electrical
Input Power
90 to 265 V AC, at 50/60 Hz
Output Power
24 V DC, regulated at 1.75A maximum
HV Cable
2 m, 30.8 pF per foot
Supplied Accessories
Operators’ Manual
PN: 23447
AC Adapter
PN: L5215R
HV Output Connector
(SHV Mating
Connector)
PN: 43874R
Optional
Accessories
None
Note
The output cable supplied with this instrument uses a coaxial cable that
has 30.8 pF/ft of capacitance at 1 kHz nominal. This cable capacitance
must be factored in as a portion of the load and will reduce slew rates
and large signal bandwidth. In applications that require maximum
performance it is suggested that the supplied high voltage coaxial cable
be kept as short as possible to reduce capacitance. Another option is
to cut the coaxial cable short and add two break out leads (1 for shield
[ground] and 1 for the center conductor) for the connection to load.
Model 2220 Specifications
Performance
Output Voltage
Range
0 to ±2 kV DC or peak AC
Output Current
Range
0 to ±10 mA DC or ±20 mA peak (for 5 ms
minimum)
Input Voltage Range
0 to ±10 V DC or peak AC
Input Impedance
10 kW, nominal
DC Voltage Gain
200 V/V
DC Voltage Gain
Accuracy
Better than 0.5% of full scale
Offset Voltage
Less than 1 V
Output Noise
Less than 50 mV rms*
Slew Rate
(10% to 90%, typical)
Greater than 100 V/µs
Large Signal
Bandwidth (-3 dB)
DC to greater than 7.5 kHz
Small Signal
Bandwidth (-3dB)
DC to greater than 50 kHz
Settling Time to 1%
Less than 50 µs for 0 to 2 kV step
Internal Capacitance
(HV Output)
300 pF
Automatic Power
Limit
Limits internal power dissipation for protection
from overheating
Stability
Drift with Time
Less than 300 ppm/hr, noncumulative
Drift with Temp
Less than 180 ppm/°C
Voltage Monitor
Noise
5 mV rms
Ratio
1/200th of the high voltage output
Current Monitor
Ratio
0.4 V/mA
DC Offset Adjust
Better than 2% of full scale
*Measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter
Copyright © 2013 TREK, INC. All specifications are subject to change. 1317/JRB