Measurement error during emptying – VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) 4 … 20 mA_HART - two-wire, approval according to LPR radio standard User Manual
Page 62
Fault description
Error pattern
Cause
Rectification
8. Measured value
jumps towards 100 %
during filling
Level
time
0
l Due to strong turbulence and
foam generation during filling,
the amplitude of the product
echo sinks. Measured value
jumps to the false echo
l Carry out false signal suppres-
sion
9. Measured value
jumps sporadically to
100 % during filling
Level
time
0
l Varying condensation or con-
tamination on the antenna
l Carry out a false signal sup-
pression or increase false sig-
nal suppression with
condensation/contamination in
the close range by editing
10. Measured value
jumps to ≥ 100 % or 0 m
distance
Le
ve
l
time
0
l Level echo is no longer de-
tected in the close range due
to foam generation or false
echoes in the close range. The
sensor goes into overfill pro-
tection mode. The max. level
(0 m distance) as well as the
status message "Overfill pro-
tection" are outputted.
l Check measuring site: Antenna
must protrude out of the socket
l Remove pollution on the an-
tenna
l Use a sensor with a more
suitable antenna
Measurement error during emptying
Fault description
Error pattern
Cause
Rectification
11. Measured value
remains unchanged in
the close range during
emptying
Level
time
0
l False echo larger than the
level echo
l Level echo too small
l Remove false echoes in the
close range. Check: Antenna
must protrude out of the socket
l Remove pollution on the an-
tenna
l Minimize interfering installa-
tions in the close range by
changing the polarization di-
rection
l After removing the false
echoes, the false signal sup-
pression must be deleted.
Carry out a new false signal
suppression
12. Measured value
jumps towards 0 % du-
ring emptying
Level
time
0
l The tank floor echo is larger
than the product echo, e.g.
with products with ε
r
<
2.5 oil-
based, solvents
l Application parameter "Me-
dium type", check Vessel
height and Floor form, adapt if
necessary
13. Measured value
jumps sporadically to-
wards 100 % during
emptying
Level
time
0
l Varying condensation or con-
tamination on the antenna
l Carry out false signal suppres-
sion or increase false signal
suppression in the close range
by editing
l With bulk solids use radar
sensor with purging air con-
nection
62
VEGAPULS
62 • 4 … 20 mA/HART - two-wire, approval according to LPR radio standard