The position of the polarisation plane is marked on the process fitting
of the instrument.

1

1

Fig. 3: Orientation of the polarisation plane with threaded version

1

Marking

1

Fig. 4: Orientation of the polarisation plane with flange version

1

Marking

When mounting the sensor, keep a distance of at least 200 mm
(7.874 in) from the vessel wall. If the sensor is installed in the center of
dished or round vessel tops, multiple echoes can arise. However,
these can be suppressed by an appropriate adjustment (see chapter
"S

etup").

If you cannot keep this distance you should carry out a false echo
storage before setup. This applies mainly if buildup on the vessel wall
is expected. In this case, we recommend repeating a false echo
storage later with existing buildup.

Mounting position

14

VEGAPULS

62 • 4 … 20 mA/HART - two-wire, approval according to LPR radio standard

4 Mounting