8 Maintenance and fault rectification

PLICSRADIO C62 • Wireless processing unit for PLICSRADIO T62

32627-EN-130701

Error code Cause

Rectification

E003

CRC error (error with

self-check)

– Carry out a reset

– Send instrument for repair

E008

PLICSRADIO T61/

T62/R62 not found

– Check connection of T61/T62/R62

– Check address of PLICSRADIO T61/

T62

– Check radio parameters and signal

strength

– R62 still in boot phase

E013

PLICSRADIO T61/

T62 signals failure,

no valid measured

value

– Check connection of the sensor

– Check sensor parameter adjustment

– Send sensor for repair

E030

Sensor in boot phase
Measured value not

valid

– Check sensor parameter adjustment

E034

EEPROM CRC error

– Switch the instrument off and on

– Carry out a reset

– Send instrument for repair

E035

ROM CRC error

– Switch the instrument off and on

– Carry out a reset

– Send instrument for repair

E036

Instrument software

not executable (dur-

ing software update

or after failed update)

– Wait until software update is finished

– Carry out another software update

E053

Sensor measuring

range not read cor-

rectly

– Communication error: Check sensor

cable and screening

Depending on the reason for the fault and the measures taken, the

steps described in chapter "Set up" may have to be carried out again.

8.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our Internet homepage

www.vega.com under: "Downloads - Forms and certificates - Repair

form".
By doing this you help us carry out the repair quickly and without hav-

ing to call back for needed information.

Print and fill out one form per instrument

Clean the instrument and pack it damage-proof

Attach the completed form and, if need be, also a safety data

sheet outside on the packaging

Please ask the agency serving you for the address of your return

shipment. You can find the respective contact data on our website

www.vega.com under: "Company - VEGA worldwide"

Reaction after fault recti-

fication