6 Setup with the integrated display and adjustment unit
PLICSRADIO C62 • Wireless processing unit for PLICSRADIO T62
32627-EN-130701
By pushing [OK] you move from the measured value indication to the
main menu.
The main menu is divided into six areas with the following functions:
•
Device settings: Includes the device-TAG, settings for network
connection such as date/time setting, …
•
Measurement loop: Includes settings for the input selection, the
outputs as well as the measurement loop TAGs
•
Display: Includes settings to the displayed measured value
•
Diagnosis Includes information on the device status and error
messages
•
Service Includes simulation, reset, PIN and language setting
•
Info: Shows serial number, software version, last change, instru-
ment features, MAC addr., …
→
Select the menu item "Device settings" with [->] and confirm with
[OK].
You can assign an unambiguous name to PLICSRADIO C62 via the
Device-TAG. This function is recommended when several instruments
are implemented and a good documentation of larger systems is
required.
→
Enter the requested values via the appropriate keys and save
your settings with [OK].
For instruments with integrated Ethernet interface, the automatic
addressing via DHCP is preset, i.e. the IP address must be assigned
by a DHCP server. Generally the instrument is contacted via the Host
name. By default, the host name consists of the serial number plus
"VEGA-" in front. As an alternative, it is also possible to enter a static
IP addr. with Subnet mask and optional Gateway addr.
Note:
Keep in mind that your modification will be only effective after a restart
of PLICSRADIO C62. You can find further information of these net-
work parameters in the supplementary instructions "RS232/Ethernet
connection" and in the Online help of the respective DTM.
Main menu
Device settings - Device-
TAG
Device settings - Host
Name/IP addr.