VEGA VEGABAR 64 Profibus PA User Manual
Page 57
Process conditions
The specifications to the pressure stage and the product temperature are used as an overview.
The specifications of the type label are applicable.
Pressure stage, process fitting
-
Thread 316L
PN
60
-
Thread Aluminium
PN
25
-
Thread PVDF
PN
10
-
Hygienic fittings 316L
PN
6, PN 10, PN 25, PN 40
-
Flange 316L
PN
16, PN 40,150 lbs, 300 lbs, 600 lbs
-
Flange with extension 316L
without PN specification, PN 16, PN 40 or 150 lbs,
300 lbs, 600 lbs
-
Flange PVDF
PN
16
Product temperature standard version, depending on the meas. cell seal
12)
-
FKM (VP2/A)
-20 … +120 °C (-4 … +248 °F)
-
EPDM (A+P 75.5/KW75F)
-40 … +120 °C (-40 … +248 °F), 1 h: 140 °C/284 °F
cleaning temperature
-
FFKM (Kalrez 6375)
-10 … +120 °C (+14 … +248 °F)
-
FFKM (Chemraz 535)
-30 … +120 °C (-22 … +248 °F)
Product temperature version with extended temperature range, depending on the meas. cell seal
as well as order specification
-
FKM (VP2/A)
-20 … +150 °C (-4 … +302 °F)
-
EPDM (A+P 75.5/KW75F)
-40 … +150 °C (-40 … +302 °F)
-
FFKM (Kalrez 6375)
-10 … +150 °C (+14 … +302 °F)
-
FFKM (Chemraz 535)
-30 … +150 °C (-22 … +302 °F)
Vibration resistance
mechanical vibrations with 4 g and 5 … 100 Hz
13)
Shock resistance
Acceleration 100 g/6 ms
14)
12)
With process fitting PVDF, max. 100 °C (212 °F).
13)
Tested according to the regulations of German Lloyd, GL directive 2.
14)
Tested according to EN 60068-2-27.
VEGABAR
64 • Profibus PA
57
10 Supplement
28252
-
EN
-081209