VEGA VEGABAR 41 HART User Manual
Page 28
28
VEGABAR 41 (HART
®
)
In the window "Span adjustment“ you first
choose in which unit the adjustment should
be carried out (mbar, psi, kPa). Please state
if you want to edit the percentage value or the
current value. If you want to edit the percent-
age value, enter the value (e.g. 80 %) in the
field. Then click to
Save
. The span adjustment
is carried out.
Setup
Sensor parameter adjustment –
Linearisation curves
Click in the window "Instrument data param-
eter adjustment“ to
Conditioning
.
Click in the window "Conditioning“ to
Lineari-
sation
.
In the window "Linearisation“ you can choose
the linearisation curves of a horizontal cylin-
drical tank or a spherical tank. When choos-
ing a user-programmable curve, you can
then click to
Edit
and start the program "Tank
Calculation“. With this program you can cal-
culate the curves of different tank shapes
(see manual "VEGA Visual Operating“). After
selection of the curve, click to
OK
.
Sensor parameter adjustment –
Integration time
Click in the window "Instrument data param-
eter adjustment“ to
Conditioning
.
Click in the window "Conditioning“ to
Integra-
tion time
.