2 function diagram, Function diagram and pa parameters – VEGA VEGASON 51P…53P Profibus PA User Manual

Page 78

background image

78

VEGASON 51P … 53P

Function diagram and PA parameters

Filter

Temperature (Float)
[temperature unit (˚C, K, ...)]

Max temperature, min temperature
(Float) [temperature unit (˚C, K, ...)]

Cal point hi, cal point lo (Float)
[sensor unit (m,ft ...)]

Level hi, level lo (Float)
[level unit (%,m,ft ...)] (*1)

Cal type (Unsigned8)
Lifeabgleich (min-max) und
Trockenabgleich möglich

Level hi

Level lo

Cal point hi

Cal point lo

Cal

Simulate value

off

on

Simulation

Zero adjust

Zero offset

Sensor offset (Float)
[sensor unit (m,ft ...)]

F Time (Float)
(sec)
(not available over
PA Parameter)

Sensor value (Float)
[sensor unit (m,ft ...)]

min

max

Max min memory

Raw-Distance
(value, status)
[in m]

Raw-Temperature
(value, status)
[in K]

Simulate procent value (DS_50)
[level unit (%,m,ft...)] (*1)
(manuf. spec. parameter)

Level offset (Float)
[level unit (%,m,ft...)] (*1)

Offset

Offset

Level (Float)
[level unit (%,m,ft...)] (*1)

8.2 Function diagram

Live adjustment (min-max) and
dry adjustment possible