• Save the adjustments with "

OK“ and you

are again in the menu window "

Meas. envi-

ronment“.

Meas. environment/Meas. conditions

• Click in the menu window "

Meas. environ-

ment“ to "Meas. conditions“.

• In the menu window "

Measuring conditions“

you click on the options corresponding to
your application.

• Confirm with

"OK“.

After a few seconds, during which the adjust-
ments are permanently saved in the sensor,
you are again in the window "

Meas. environ-

ment“.

Meas. conditions/Sonic velocity

In the menu item "

Sonic velocity“ adjustments

are only necessary when using the sensor in
a gas composition deviating from air. If you
measure in a gas composition deviating from
air, e.g. pure nitrogen, the sonic velocity
changes. Without correction, the measuring
result would be incorrect.

In the menu "

Sonic velocity“ the sonic velocity

can be set to a value corresponding to a
different gas composition.

• With "

OK“ you save the adjustments in the

sensor.

• Click in the window "

Meas. environment“ to

"

Quit“.

You are again in the menu window "

Sensor

optimisation“.