Pa-bus, Adjustment with adjustment module minicom, Product description – VEGA VEGASON 51P…53P Profibus PA User Manual
Page 10
10
VEGASON 51P … 53P
Product description
Adjustment with the SIMATIC PDM ad-
justment program
To adjust all essential functions of the VEGA
sensor with the adjustment station SIMATIC
PDM from Siemens, a so-called EDD is re-
quired. Without this EDD, only the basic func-
tions, such as min./max. live adjustment or
the integration time, can be adjusted with the
PDM adjustment program. Further important
adjustment functions, such as the input of the
"
Meas. environment“, a dry adjustment or a
false echo storage are not available without
EDD. After integration of the EDD files into the
Simatic PDM adjustment software, all impor-
tant adjustment functions are accessible. If
the file is not on hand, the obligatory GSD
(instrument master file) as well as the EDD
(Electronic Device Description) necessary for
PDM can be downloaded from the VEGA
Homepage (http://www.vega.com).
-
+
ESC
OK
2
4
Tank 1
m (d)
12.345
-
+
ESC
OK
Tank 1
m (d)
12.345
PA-Bus
Adjustment with detachable adjustment module. The
adjustment module can be plugged into the ultrasonic
sensor or the external indicating instrument VEGADIS
50.
Detachable adjustment module MINICOM
-
+
ESC
OK
Tank 1
m (d)
12.345
Adjustment with adjustment module
MINICOM
With the small (3.2 cm x 6.7 cm) 6-key ad-
justment module with display in the sensor,
the sensor-relevant adjustments can be car-
ried out directly on the sensor.
The adjustment module can be plugged into
the ultrasonic sensor or into the optional,
external indicating instrument.
With the adjustment module MINICOM, the
sensor can be adjusted to the actual measur-
ing conditions, and the basic functions can
be set. In addition to the measuring condi-
tions and simulation mode, the Profibus ad-
dress can be adjusted and a false echo
storage can be carried out, see "6.3 Adjust-
ment with MINICOM“.
max. 2.5 m