44

VEGASON 54K … 56K

Set-up - Adjustment with the adjustment module MINICOM

Menu schematic for the adjustment module MINICOM

Sensor
opti-
mise

PPara-
meter

Sensor

m(d)

4.700

Meas.
enviro
nment

Opera-
ting
range

Begin

m (d)

0.50

End

m (d)

6.00

Meas.
condit
ions

Measur
ing in
gas

Correc-
tion
Now

!

OK

?

Meas.
dist.

mm (d)

2,381

SON
51K

2.00

When switching on, the sensor
type and the software version are
displayed for a few seconds.

2.

5.

1.

Adjust
ment

Signal
condit
ioning

Out-
puts

Curr.
out-
put

Sensor
dis-
play

Error
mode

22mA

Prop.
to

Dis-
tance

Curr.
out-
put

4-20mA

w.out
medium

Adjust
ment
in

m(d)

with
medium

0.0 %

at

m (d)

XX.XXX

100.0%

at

m (d)

XX.XXX

Min.
adjust
at %

XXX.X

Max.
adjust
at %

XXX.X

Sca-
ing

Lin.
curve

Linear

Integr
ation
time

0 s

0 %
corres
ponds

XXXX

100 %
corres
ponds

XXXX

Dezi-
mal
point

888.8

Prop.
to

Mass

Unit

Kg

3.

4.

8.

Correc-
tion
Now

!

OK

?

Sound
speed

m/s

333

Confi-
gura-
tion

Sensor
Tag

Sensor

Sensor
addr.

0

Meas.
unit

m (d)

Multidrop operation (HART

®

sensor ad-

dress):
• Sensor address zero: The sensor outputs

beside the 4…20 mA signal also a digital
(HART

®

) level signal.

• Sensor address 1…15: the sensor delivers

only a digital (HART

®

) level signal. The

sensor current is frozen to 4 mA (power
supply).

Condit
ions

Liquid

Agitat
ed sur
face

No

Foam-
ing
prod.

No

Fast
change

No

Measur
ing in
gas

No

Mul-
tiple
echoes

No

Condit
ion

Solid

High
dust
level

No

Large
angle
repose

No

Fast
change

No

Measur
ing in
gas

No

Mul-
tiple
echoes

No