VEGASON 54K … 56K

41

Adjustment without medium

(adjustment independent of the actual level)

Key adjustment

Display indication

Set-up

2. Operating range

Without special adjustment, the operating
range corresponds to the measuring range.
The measuring range has already been ad-
justed with the min./max. adjustment. Gener-
ally it is useful to set the operatign range
slightly larger (approx. 5 %) than the measur-
ing range.

Example:
Min./max. adjustment: 0.300 … 5.850 m;
adjust operating range to approx.
0.250 … 6.000 m.

3. Adjustment

Under the menu item "

Adjustment“ you inform

the sensor which measuring range it should
operate with.

You can carry out the adjustment with or
without medium. Generally you will carry out
the adjustment without medium, as you can
then adjust without a filling cycle.

Adjust-
ment

Para-
meter

Sensor

m(d)

4.700

w.o
medium

Ad-
just-
ment
in

m(d)

(min. adjustment)

OK

OK

OK

OK

Max.

Min.

100 % (1.270 m) correspond
to 1200 liters

Span (4.58 m)

0 % (5.850 m) corresponds
to 45 liters

+

OK

Ad-
just-
ment
in

m(d)

0.0%

at

m (d)

XX.XXX

OK

+

or

The distance indication flashes
and you can choose "feet“ and
"m“.

Confirm the adjustment with
"

OK“.

With"

+“ and "–“ you adjust the

percentage value for the min.
value or the lower level (exam-
ple 0.0 %).

The adjusted percentage
value is written in the sensor
and the distance corres-
ponding to the percentage
value flashes.