beautypg.com

VEGA VEGASON 54K…56K User Manual

Page 14

background image

14

VEGASON 54K … 56K

Technical data

Ambient conditions

Ambient temperature (housing)

-20°C … +60°C

Process temperature (transducer)

- VEGASON 54, 55

-40°C … +80°C (StEx: -20°C … +75°C)

- VEGASON 56

-40°C … +150°C

- storage and transport temperature

-40°C … +80°C

Vessel pressure max. (gauge pressure)
- VEGASON 54

Version A

0.5 bar (flange version)

Version B … C

0.5 bar

Version D

0.5 bar

- VEGASON 55

Version A

0.5 bar (flange version)

Version B … C

0.5 bar

Version D

0.5 bar

- VEGASON 56

Version A

0.5 bar (flange version)

Version B … C

0.5 bar

Version D

2 bar

Protection

- sensor

IP 67

- transducer, process

IP 68

Protection class

- two-wire sensor

II

- four-wire sensor

I

Overvoltage category

III

Self-heating
at 40°C ambient temperature
- sensor

45°C

- transducer, process

55°C

Ex technical data

Classification

m (encapsulation of the transducer)

Temperature class (permissible ambient
temperature around the transducer when
used in Ex areas)

- T6

42°C

- T5

58°C

- T4

60°C

- T3

60°C

Ex approved in category or zone

- VEGASON 56

ATEX

Zone 1 (II 2G)

IEC, CENELEC, PTB

Zone 1

- VEGASON 54 … 56

ATEX

Zone 21/22 (II 2D/3D)

IEC, CENELEC, DMT

Zone 10/11