beautypg.com

VEGA VEGAFLEX 54P (Profibus PA) User Manual

Page 45

background image

VEGAFLEX 54P (Profibus PA)

45

Technical data

Connection cables

Two-wire sensor
- power supply and signal via one two-wire cable.

The line resistance depends on the supply voltage (see diagram).

Cross-section area of conductor

general 2.5 mm

2

Ground connection

max. 4 mm

2

Cable entry

2 x M20 x 1.5 (cable diameter 5 … 9 mm)

2 x

1

/

2

“ NPT with ExD housing

CE conformity

VEGAFLEX sensors meet the protective regulations of EMC (89/336/EWG) and NSR
(73/23/EWG). Conformity has been judged acc. to the following standards:
EMC

Emission

EN 50 081 - 1: 1992

Susceptibility

EN 50 082 - 2: 1995

NSR

EN 61 010 - 1: 1993

Display

Indication

scalable analogue and digital measured value
display (option).

An external measured value display powered by the sensor can be mounted up to 25 m
away from the sensor.

Signal output

Signal output

digital output signal in two-wire technology
(Profibus)

Two-wire technology:
The digital Profibus output signal (measuring signal) is modulated onto the power supply
and is further processed in the signal conditioning instrument or in the processing system.