beautypg.com

VEGA VEGAFLEX 54P (Profibus PA) User Manual

Page 44

background image

44

VEGAFLEX 54P (Profibus PA)

Technical data

Characteristics

Min. span between full and empty adjustment

- digital output signal

10 mm

Adjustment time

1)

> 1 s (depending on parameter setting)

Ambient conditions

Vessel pressure

-1 … 40 bar

Ambient temperature on the housing

-40°C … +60°C

Process temperature

-40°C … +150°C

Storage and transport temperature

-40°C … +80°C

Protection

IP 66/IP 67 (meets both protection classes)

Protection class
- two-wire sensor

II

- four-wire sensor

I

Overvoltage category

III

Max. tensile load

cable ø

4 mm: 1 KN

rod ø

8 mm: 1 KN

Min. dielectric constant

ε

r

> 1.7

Process fittings

VEGAFLEX 54

G1 A, 1“ NPT, G 1

1

/

2

A, 1

1

/

2

“ NPT

of 1.4435 (stst; 316L) or
Hastelloy C22 plated

Ex-technical data

Ex-Zone 0 acc. to ATEX II 1 G or II 1/2 G or II 2 G EEx ia II C T6, ATEX II 1/2 G or II 2 G
EEx d ia II C T6 and ATEX II 1/2 G
The permissible operating data of VEGAFLEX sensors for Ex areas are stated in the certifi-
cate.

Materials

Housing

PBT (Valox) or
Aluminium (powder coated)

Rod

1.4435 (stst; 316L) or Hastelloy

Cable

1.4401 (stst; 316L)
gravity weight 1.4435

2)

The adjustment time (also actuating, response time or adjustment period) is the time required by the sensor
to output the level correctly (with max. 10 % deviation) after a sudden level change.