Ranger™ 3000, Series en-15 – Ohaus RANGER 3000 COMPACT SCALES Manual User Manual
Page 17
Ranger™ 3000
Series EN-15
Zero Range
[ZErO]
Set the percentage of scale capacity that may be zeroed.
2%
= zero up to 2 percent of capacity
10% = zero up to 10 percent of capacity
Auto Tare
[A.tArE]
Set the automatic tare functionality.
OFF
= Automatic Tare is disabled
ON
= the first stable gross weight is tared
ON-ACC = stable gross loads within the accept limits are tared (in Check weighing mode)
Beeper Signal
[Bp.sIG]
Set how the beeper responds in Check Weighing mode.
OFF
= the beeper is disabled
ACCEPT = the beeper will sound when the weight is within the Accept range.
UNDER = the beeper will sound when the weight is below the Under setting.
OVER = the beeper will sound when the weight is above the Over setting.
OVER- = the beeper will sound when the weight is below the Under setting or above the
UNDER Over setting.
Beeper Key
[BP.key]
Set whether the beeper sounds when a button is pressed.
OFF
= no sound
ON
= sound
Accumulation
[aCCUmM]
Set the accumulation functionality.
OFF
= accumulation disabled
AUTO = automatic accumulation
MAN
= manual accumulation
End Setup
[End]
Advance to the next menu or return to the top of the current menu.
4.4 Readout Menu
Enter this menu to set user preferences. Default settings are in bold.
Reset:
no
, yes
Stable Range: 0.5, 1, 2, 5
Filter:
low, medium, high
Auto-Zero
Tracking:
off, 0.5, 1, 3
Light:
off, on, auto
Auto off:
off
, 1, 5, 10
End Readout: Exit menu