Series – Ohaus RANGER 3000 COMPACT SCALES Manual User Manual
Page 16

EN-14
Ranger™ 3000
Series
Span
[SpaN]
Initiates a span calibration procedure (zero and span)
Lin
[LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
GEO
[GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current
location. Settings from 0 to 31 are available with 12 being the default.
Refer to table 4-2 to determine the GEO factor that corresponds to your location.
End Cal
[End]
Advance to the next menu or return to the top of the current menu.
4.3 Setup Menu
Enter this menu to set scale parameters. Default settings are in bold.
Reset
[reset]
Reset the Setup menu to factory defaults.
NO
= not reset
YES = reset
Power on unit
[pwWr.UN]
Set the unit of measure displayed at startup.
AUTO = last unit in use when turned off
kg
= kilograms
g
= grams
lb
= pounds
oz
= ounces
lb:oz = pound ounces
Span:
Perform
Linearity:
Perform
Geographic
Adjustment:
Set 0…12…31
*
End Calibration: Exit menu
*Bold always represents factory default value.
Reset:
no
, yes
Power on unit: auto, kg, g, lb, oz, lb:oz
Zero Range:
2%, 10%
Auto Tare:
off
, on, on-acc
Beeper Signal: off, accept, under, over, under-over
Beeper Key:
off
, on
Accumulation: off, auto, manual
End Setup:
Exit menu