Settings in the menu

4.6.3

COM1/COM2 -> PARAMET – communication parameter

4.6.4

COM1/COM2 -> RESET COM1/RESET COM2 – resetting serial interface to factory
settings

4.6.5

OPTION – configuring options
If no option is installed or is not yet configured, N.A. appears in the display.

bAUd

Selecting baud rate: 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 baud

PAritY

Selecting parity: 7 none, 8 none, 7 odd, 8 odd, 7 even, 8 even

H.SHAKE

Selecting Handshake: NO, XONXOFF, nEt422, nEt485 (network operation
as per RS485 standard via the optional RS422/RS485 interface, only for COM1)

NET.Addr

Assigning network address: 0 ... 31, only for NET 485

ChECSuM

Activating checksum byte (appears only in TOLEDO Continuous mode)

Vcc

Switching 5V voltage, e.g. for a bar code reader, on / off

SUrE?

Confirmation inquiry

Reset interface settings to factory settings with

Yes

Do not reset the interface settings with

No

EtH.NEt

Configuration of the Ethernet interface

IP.AddrS

Enter IP address

SUBNEt

Enter Subnet address

GAtEWAY

Enter Gateway address

USb

Configuration of the USB interface

USb TEST

Test of the USB interface. After the test has been passed, rEAdY appears in the
display.

diGitAL

Configuration of the digital inputs/outputs

IN 1 ... 4

Configuring inputs 1 ... 4

OFF

Input not assigned

ZErO

ZERO Key

tArE

TARE Key

PriNt

PRINT Key

rEF 10

Sample 10 Key

rEF n

Sample Size Key

SCALE

Scale Select Key

UNIt

Count Weight Key