beautypg.com

4 tare - tare function, 5 zero - automatic zero update – Ohaus RANGER COMPACT COUNTING SCALES Manual en User Manual

Page 27

background image

Settings in the menu

Ranger RC

27

4.3.4

TARE – tare function

4.3.5

ZERO – automatic zero update

4.3.6

RESTART – automatic saving of zero point and tare value

4.3.7

FILTER – adaptation to the ambient conditions and the weighing type

A-tArE

Switching on/off automatic taring

CHAIn.tr

Switching on/off chain tare

A.CL-tr

Switching on/off automatic taring with automatic clearing of the tare weight when the
load is removed from scale

AZM

On certified scales, this menu item does not appear.
Switching on/off automatic zero update and selecting zeroing range.
Possible settings: OFF; 0.5 d; 1 d; 2 d; 5 d; 10 d

ON/OFF

When the Restart function is activated, the last zero point and tare value are saved.
After switching off / on or after a power interruption, the device continues to work with
the saved zero point and tare value.

VIbrAt

Adaptation to the ambient conditions

LOW

Very steady and stable environment. The scale works very quickly, but is very
sensitive to external influences.

MEd

Normal environment. The scale operates at medium speed.

HIGH

Restless environment. The scale works more slowly, but is insensitive to external
influences.

PrOCESS

Adaptation to the weighing process

UNIVEr

Universal setting for all weighing samples and normal weighing goods

dOSING

Dispensing liquid or powdery weighing samples

StAbILI

Adjusting the weighing speed

FASt

The scale operates very fast.

StAndrd

The scale operates at medium speed.

PrECISE

The scale operates with the greatest possible reproducibility.

The slower the scale works, the greater the reproducibility of the weighing results.