18
Trooper Count Scale
4.5 Setup Menu
Procedure
With the Scale ON, press and hold theSAMPLE Menu
button until CAL MENU is displayed.
NOTE: The calibration menu is covered in Section 5.
Press M- button to advance to SETUP MENU, then
press Enter button. Navigate through the menu as
explained in section 4.3.1.
Press Enter button to get into the RESET sub-menu.
YES will change all menu settings to their default set-
tings. Library data will not be affected.
ON will set the Scale to LFT configuration. In the ON
setting, the zero setting will be limited to +/- 2%. In
the OFF setting, the zero setting is 100%. Also, in the
ON setting, the LFT readability is set. In the OFF posi-
tion, the readability is set to the Enhanced value.
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
BLINKING
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
BLINKING
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
Three setting are available, ALL, SIGNAL and OFF.
ALL = the beeper sounds when buttons are pressed or
when error or target signals are needed. SIGNAL =
beeps on errors (three beeps) or at count target (con-
tinuous). OFF = is always off.
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
kg
g
lb
oz
Net
B/G %Ac
T
PT APW MR
M
<
>
PCS
WEIGHT
DATA
COUNT
Selected calibration units are either lb or 1000g (kg).
A selection of 1/3, 2/3 or full span calibration values
are available. The numeric values of the calibration
points are displayed for the specific Scale capacity
and calibration unit.
Backlight has three available settings; ON, OFF, AUTO.
ON = backlight is always on. AUTO = backlight turns
on with any key press or a weight change (5d) and
remains on for 5 seconds. OFF = backlight is always
off. OFF or AUTO is recommended for battery opera-
tion.
ON = Scale turns off in 5 minutes if no key press or
weight changes (5d) are detected. OFF = auto off
feature is disabled. ON is recommended for battery
operation.