beautypg.com

Trooper scale – Ohaus TROOPER COMPACT SCALES Manual multi User Manual

Page 38

background image

EN-32

Trooper Scale

6.2 Error Codes List

The following list describes the various error codes which can appear on the display.

LoBat

Is indicated when batteries are weak. Approximately 20 minutes of operating time remain.

Error 1

Indicates an overload condition.

Error 2

Indicates an underload condition.

Error 7

EEPROM data incorrect.

Error 14

Zero exceeds ZERO% and cannot be zeroed.

Err 21

Calibration data does not match current full scale, Grad and Cal Point settings. Settings must
be restored or the Scale must be recalibrated using the current settings.

6.3 Service Information

If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus
Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-free at (800)
526-0659. An Ohaus Product Service Specialist will be available to help you.

6.4 Accessories

Description

Ohaus Part No.

RS232 Interface Cable/SF42 Printer

80500573

RS232 Interface Cable/PC 25 Pin

80500431

RS232 Interface Cable/PC 9 Pin

80500433

Printer

SF42

6.5 Technical Data

Materials

ABS Housing

Keypad/display overlay polyester