beautypg.com

Trooper scale 6 care and maintenance (cont.) – Ohaus TROOPER COMPACT SCALES Manual multi User Manual

Page 37

background image

EN-31

Trooper Scale

6 CARE AND MAINTENANCE (Cont.)

Cannot display weight in desired
weighing unit.

RS232 not working.

Unable to calibrate unit.

Desired unit not set to ON in Readout
menu.

RS232 communication parameters set up
incorrectly.

Improper or loose cable connections.

Software Lockout switch set to ON and
Lock Switch on the circuit board set to
open position.

Incorrect value for calibration mass.

Enable desired unit in Readout
menu. See paragraph 4.4.4

Conversion too large (typically in
g).

Verify communication param-
eters.

Check cable connections.

Set LCL to OFF in the LocSW
menu, and set Lock Switch on the
circuit board to ON position. Re-
fer to paragraph 4.4.6.

Use correct calibration mass.

SYMPTOM

PROBABLE CAUSE(S)

REMEDY

6.1 Troubleshooting (Cont.)