beautypg.com

Trooper scale, Calibration and sealing (cont.) – Ohaus TROOPER COMPACT SCALES Manual multi User Manual

Page 29

background image

EN-23

Trooper Scale

Cal in kg:

Span cal choices

Linearity cal (fixed)

TR3RS

1, 2, 3kg

2 & 3 kg

TR6RS

2, 4, 6kg

4 & 6 kg

TR15RS

5, 10, 15kg

10 & 15 kg

TR30RS

10, 20, 30kg

20 & 30 kg

Cal in lb:

Span cal choices

Linearity cal (fixed)

TR3RS

2, 4, 6 lb

4 & 6 lb

TR6RS

5, 10, 15 lb

10 & 15 lb

TR15RS

10, 20, 30 lb

20 & 30 lb

TR30RS

20, 40, 60 lb

40 & 60 lb

5. CALIBRATION AND SEALING (Cont.)

Procedure (Cont.)
TURNING SOFTWARE LOCK OFF

For span calibration, there are a number of choices for
calibration. The highlighted number is the default.

To change the span calibration value read the setup sec-
tion of the menu on page 10. In the Setup menu, change
the CP selection to the desired value.

When the desired selection has been chosen and the
masses are available, you are ready to begin the calibra-
tion routine.

Procedure

SPAN CALIBRATION

With the Scale ON, press and hold the G/N/T/MENU
button until MENU is displayed. When you release t h e
G/N/T/

MENU button, CAL is displayed. If the display

shows SETuP, read the section on turning off the Calibra-
tion Lock Switch or check the settings for legal for trade.

Press G/N/T/

MENU button, SPAN Y is displayed.

Press G/N/T/

MENU button, -C- is displayed. The scale

MUST be stable during this period and is establishing a
zero point. After a few seconds, the requested weight
value is displayed.