Trooper Scale

4.4.3 Setup Menu

The Setup Menu is used to set up the Scale for the first time.

Procedure

With the Scale ON, press and hold the G/N/T/MENU
button until MENU is displayed. When you release G/N/
T/

MENU button, SETUP is displayed when the CALIBRA-

TION Lock Switch (software) is in the locked position.
When the CALIBRATION Lock Switch is locked, the Scale
will not permit calibration.

Press PRINT/UNITS button, SETuP is displayed.

Press G/N/T/

MENU button, LFTOFF is displayed. Legal

for trade selections are:

'LFT ON' - LFT is ON
'LFTOFF' - LFT is OFF
'LFTCAn' - LFT is set for Canada

Press PRINT/

UNITS button and select either ON, OFF or

Canada.

Press G/N/T/

MENU button, 0 18 is displayed. This isthe

Zero 2%, 18% or 100% setting. 2% - zero operation
range is - 2% to + 2%. 18% - zero operating range is -
2% to +18%, 100% - zero operation range is -2% to
+100%.
NOTE:

: :

:

: If LFT is ON, only 2% and 18% are available.

Press PRINT/

UNITS button, and select either 2%, 18%

or 100%.

Press G/N/T/

MENU button, CAL Un kg is displayed.

This is the calibration unit setting. Selections are:

'lb' - calibration unit is lb

'kg' - calibration unit is kg.

Press PRINT/

UNITS button, and select either kg or lb.

NOTE: The Scale is factory calibrated. After changing
the calibration unit, the Scale must be recalibrated
befoe using!