beautypg.com

Trooper scale – Ohaus TROOPER COMPACT SCALES Manual multi User Manual

Page 24

background image

EN-18

Trooper Scale

4.4.5 Print Menu

The Print menu provides data communication settings which can be entered. It contains 9 submenus: Reset, Baud rate,
Parity Bit, Data Length, Stop Bits, Auto Print, Interval, Stable and Save.

Procedure

PRINT
To select any of the items in the Print menu, proceed as
follows: NOTE: If you have entered from the preceeding
menu, disregard the first step.

With the Scale ON, press and hold the G/N/T/MENU but-
ton until MENU is displayed. When you release the G/N/
T/

MENU button, CAL is displayed, then press PRINT/

UNITS button, until Print is displayed.

RESET

Press G/N/T/

MENU button, rESEtn is displayed. This al-

lows resetting the Print menu to factory defaults.
rESETn = no does not reset settings. rESETy = yes will
reset the entire Print menu as follows:
Baud rate =2400, parity =none, data length=7,
stop bit=2.

Press PRINT/

UNITS button, and select N or Y.

BAUD RATE

Press G/N/T/

MENU button, bd9600 is displayed.

Press PRINT/

UNITS button, and select desired baud

rate. Baud rate selections are: 1200, 2400, 4800 9600
and 19200. 9600 is the default setting.

PARITY

Press G/N/T/

MENU button, PAr NO is displayed. This is

the parity bit.

Press PRINT/

UNITS button, and select desired parity of

NO=none, Odd=odd, E=even. Default setting is none.