beautypg.com

Trooper scale, Operation (cont.) – Ohaus TROOPER COMPACT SCALES Manual multi User Manual

Page 15

background image

EN-9

Trooper Scale

NOTE: If you hold this button down too long, the display
will advance to another measuring unit.

3.8.1 RS-232 Pin Out.

The table illustrates the pin-out connections on the RS-
232 connector.

3.7 RS232 Commands

All communication is accomplished using standard ASCII format. Characters shown in the following table are acknowl-
edged by the Scale. Invalid command response "ES" error indicates the Scale has not recognized the command.
Commands sent to the Scale must be terminated with a carriage return (CR) or carriage return-line feed (CRLF). Data
output by the Scale is always terminated with a carriage return-line feed (CRLF).

3.7.1 Output Formats

Command
Character

Description

?

Print current mode: kg, g, lb., oz.

P

Same as pressing PRINT button.

T

Same as pressing TARE button.

Z

Same as pressing ZERO button.

xS

Print Stable only. Where x=0 Off, and x=1 On

AS

Automatically send data when stable after motion.

xxxxS

Send at interval. Where xxxx=1 to 3600 seconds.

CS

Send as fast as possible (continuous print)

M

Increment to next enabled unit

To turn auto printing, interval printing or continuous printing off, send P to reset normal printing mode.

RS232 USER COMMAND TABLE

3. OPERATION (Cont.)

1 N/C
2 Data In (RXD)
3 Data Out (TXD)
4 N/C
5 Ground
6, 7, 8, 9 N/C

3.8 Printing Data

Printing data to an external computer or printer requires that the communications parameters in the Print menu, be set
first. Refer to paragraph 4.4.5 Print Menu for proper set up.

To print data, press PRINT/

UNITS button with a short press. The display acknowledges by momentarily blinking off.

Data output can be initiated in one of two ways:
1. By pressing the PRINT/

Units button, or 2. Sending a print command (“P”) from a computer.

Output Formats
The output format is as follows:

Weight*

Spaces

Unit

Stable

Legend

CR

LF

Length:

9

1

3

1

1

1

1

blank=stable G,N,T
"?"= not stable

* Displayed weight sent right justified with lead zero blanking. Nine characters (fixed) include:
decimal point (1), weight (7 max), polarity (1) : blank if positive, floating negative (1)